市販試薬を用いたディスク型ガンマ線標準線源の調製法Preparation Method of Gamma-ray Disk Source Using Commercially Available Chemical Reagents

1明治大学大学院理工学研究科応用化学専攻Applied Chemistry Course, Graduate School of Science and Technology, Meiji University ◇ 214–8571 神奈川県川崎市多摩区東三田1–1–1 ◇ 1–1–1 Higashimita, Tama-ku, Kawasaki, Kanagawa Pref. 214–8571, Japan

2明治大学理工学部応用化学科Department of Applied Chemistry, School of Science and Technology, Meiji University ◇ 214–8571 神奈川県川崎市多摩区東三田1–1–1 ◇ 1–1–1 Higashimita, Tama-ku, Kawasaki, Kanagawa Pref. 214–8571, Japan

3明治大学研究・知財戦略機構Organization for the Strategic Coordination of Research and Intellectual Properties, Meiji University ◇ 214–8571 神奈川県川崎市多摩区東三田1–1–1 ◇ 1–1–1 Higashimita, Tama-ku, Kawasaki, Kanagawa Pref. 214–8571, Japan

受付日:2016年11月1日Received: November 1, 2016
受理日:2017年8月30日Accepted: August 30, 2017
発行日:2018年2月15日Published: February 15, 2018
HTMLPDFEPUB3

市販試薬中に存在する天然放射性核種を用いて,γ線スペクトロメトリー用のディスク型γ線源の作製を試みた。天然放射性核種である40Kと176Luを含む市販試薬をそれぞれ溶解後,ゼラチンと少量の食用着色料を添加し,液体をガラス繊維フィルターに滴下することで線源とした。滴下量から算出した放射能を用いて,γ線スペクトロメトリーにおけるディスク型γ線標準線源の検出効率曲線を作成した。検出効率曲線の妥当性は,ディスク型γ線標準線源と同型の認証標準試料の分析結果を認証値と比較することで評価した。認証標準試料中の放射性核種濃度の認証値と作成した検出効率曲線から得られた定量値はよく一致した。固相抽出とγ線測定を組み合わせて雨水中の放射性核種の定量分析に適用したところ,良好な定量値が得られた。ディスク型γ線標準線源は,市販の試薬から作製できるため,管理区域外でのγ線スペクトロメトリーに適用できる点が利点である。

An estimation method of detection efficiencies for γ-ray spectrometry has been developed using standard sources consisting of natural radionuclides of 40K and 176Lu included in commercially available chemical reagents. The γ-ray standard source was prepared in the shape of membrane disk by dropping aqueous solution of KCl or Lu2O3 and gelatin reagents on a glass fiber membrane. Detection efficiencies were calculated using both areas of photo peaks in γ-ray spectra and weights of the chemical reagents. Detection efficiencies in the energy range from 202 to 1461 keV can be obtained by determination of the standard source. Experimental results of activity concentrations in radioactive standard materials were in good agreement with the certified value. The present method using the γ-ray disk source was applicable to determination for short-lived radionuclides in rainwater sample collected on the chelating resin disk. The method enables to estimate activities of several radionuclides in environmental samples outside a radiation controlled area.

Key words: γ-ray standard disk source; detection efficiency; natural radionuclide; commercially available chemical reagent; outside a radiation controlled area